Statement of Results
Derived wavelength values are shown in the table
below. The rightmost column shows
absolute error for each derived wavelength, measured against the closest
published value.[1] These agree well with probable error, which
is based on uncertainty in reading the instrument scale.
Line
Wavelengths
Source |
Line Description |
m |
λ, nm |
±λ, nm |
Δλ,
nm |
|
|
|
|
probable error |
compared with published value |
|
|
|
|
|
|
H |
faint
violet |
1 |
410.8 |
0.2 |
0.6 |
|
violet |
1 |
434.3 |
0.2 |
0.3 |
|
blue-green |
1 |
486.3 |
0.2 |
0.1 |
|
red |
1 |
656.3 |
0.2 |
0.0 |
|
|
|
|
|
|
|
violet |
2 |
434.7 |
0.1 |
~ |
|
blue-green |
2 |
486.7 |
0.1 |
~ |
|
red |
2 |
656.5 |
0.1 |
~ |
|
|
|
|
|
|
He |
faint
violet |
1 |
438.8 |
0.2 |
0.0 |
|
violet |
1 |
447.4 |
0.2 |
0.2 |
|
faint blue |
1 |
471.7 |
0.2 |
0.4 |
|
blue-green |
1 |
501.7 |
0.2 |
0.1 |
|
green |
1 |
505.4 |
0.2 |
0.6 |
|
yellow |
1 |
587.8 |
0.2 |
0.3 |
|
red |
1 |
668.5 |
0.2 |
0.6 |
|
faint red |
1 |
707.5 |
0.2 |
1.0 |
Discussion of Results
For mode 2 hydrogen lines, the probable error values
came out half the size of those for of mode 1. That seemingly odd result is due
to the division by the mode# in calculating wavelength. (See Sample
Calculations). The formula for probable
error involves the measured wavelength as a factor, causing the result to be
divided by m. (A similar outcome
multiplies the calculated dispersion for each mode 2 line - see Sample
Calculations, and Data Analysis.)
Sodium reference lines at 589.0 nm and 589.6 nm,
were used to find the slit separation d
(1659.2 ± 0.1 nm). This fundamental value was used in calculating wavelength
and probable error in wavelength for non-reference lines, grating lines per cm,
grating total lines N, dispersion D, and half-width for the reference
lines.
Grating size was found to be 1.53 x 104
lines per inch (6027.0 lines per cm), within 2% of the specified 15,000 lines
per inch.
Grating width (2.2 cm) was measured with a ruler,
and used to find N, the absolute
number of slits in the grating (1.3 x
104). N was then used to
compute the half-width angle of each sodium (reference) line, from which was
derived a theoretical lower limit of resolution (0.095 nm). An upper or
practical limit for resolution is evident in the distance (0.6 nm) between the
two sodium lines, which we were able to distinguish with some care.
Spectroscope resolving power R for mode 1, is equal to N
(1.3 x 104) by its formula (Nm). R is also expressed as the ratio of wavelength midpoint to
wavelength difference, for 2 just-resolvable lines. These were our friendly
sodium lines, but R by this method
(1 x 103) was found to be misbehaving by a factor of 10. This was
designated as "unhelpful" until it was realized that the wavelength
difference used here was 0.6 nm, an upper limit for resolution. Plugging in the
theoretical lower limit of 0.095 nm, yields a value for R (6 x 103 ) that is off only by a factor of 2.
Dispersion D
is expressed as the ratio of difference in angle, to difference in wavelength
for any two lines. D can also be
calcuated for individual lines. Expressed either way, the spectroscope showed
dispersion of 60,000 to 70,000 radians per nanometer for mode 1 lines, and 2 to
3 times as much for mode 2 lines.
Scatter plots for Hydrogen and Helium lines (see
Graphs) show trends in absolute error. The graph of error vs. wavelength for
Hydrogen lines shows decreasing error for increasing wavelengths. The opposite
trend is shown on the corresponding Helium graph. This graph appears to show
two distinct upward trends - could be random scatter, or could reflect the two
of us switching off the task of reading the scale for successive lines.
Error as percentage of wavelength, vs. wavelength
(shown for Hydrogen, inspected but not shown for Helium) closely followed the
same trends in each case. This suggests that change in wavelength doesn't cause the corresponding change in error.
Significantly, measured wavelengths were almost
uniformly greater than published wavelengths. These smooth, distinct trends
indicate systematic error, likely associated with instrument alignment, or
experimental method, or even the process of assigning values from the reference
manual.